Calibration is an essential step in every precision measurement. It establishes the relationship between the sound pressure and the electrical output of the microphone.

Calibration can be performed at either 250 Hz or at 1 kHz. Both frequencies have their advantages and disadvantages.

250 Hz is in the frequency range where the frequency response of almost all microphones is flat and will therefore give the most accurate calibration.

If, however, the microphone is used in connection with measuring equipment with weighting filters, for example a sound level meter with A-weighting, the 250 Hz signal will be attenuated by the weighting network. In this case it might be advantageous to use a 1 kHz calibration frequency.

At 1 kHz, the weighting filters have 0 dB attenuation and will therefore not affect the calibration.

On the other hand, since the frequency response of the microphone is not flat, a small error could be introduced.

Products in Acoustic Transducers

Special microphones

Special microphones are often required for applications where there are particular...

view more

Power modules

Why a Power Module? An externally polarized condenser measurement microphone cannot work...

view more

Test fixtures

Including products for ear- and headphone testing, hearing-protector testing and telephone...

view more

Preamplifiers for microphone cartidges

The output from a condenser microphone is a very high-impedance signal and is therefore...

view more

Accessories

GRAS Sound & Vibration offers a wide range of standard accessories in the form of...

view more

Artificial ears, ear & mouth simulators

GRAS artificial ears and ear & mouth simulators are designed to enable measurements...

view more